General Info
Summary of system capabilities
Vendor: Allwin21 Corp.
Model: AccuThermo AW 610
Purpose: Short duration, high-temperature processing of 4" or 6" wafers
Model: AccuThermo AW 610
Purpose: Short duration, high-temperature processing of 4" or 6" wafers
Important
- Material restrictions in effect: click here
- Training required prior to use: click here
- More info for lab members: Manuals and Processes
The AW610 rapid thermal processing (RTP) system uses high intensity visible radiation to heat single wafers for short process periods of time (typically 10 minutes or less) at precisely controlled temperatures (400°C-1200°C). The system is capable of annealing substrates in N2, Ar, O2 environments. The system is a cold wall, quartz chamber, which is heated with top and bottom arrays of high intensity halogen lamps using a closed-loop temperature control using either a pyrometer or a thermocouple. Typical processes for this tool include metal-silicide formation, very thin thermal oxidation and relief of ion implantation damage.
Important Note on Cleanliness
Any preprocessed wafer exposed to organic material (photoresist, e-beam resist, wax, oil) or other material that can degrade and degas at high temperature must undergo a thorough wet or plasma cleaning before being introduced into this system.
Additional Information
- System is equipped to handle 3” – 6” wafers placed directly on the tray
- Multi-pocket susceptor is available for annealing of III-V materials or IR transparent substrates. Substrate size is limited to <2” diameter in this case.