Authorized Materials
Summary of material constraints
Lab members wishing to introduce new materials into this system must first obtain authorization from staff. New material requests must be initiated by submitting a Process Review Request form.
Non compliance will result in loss of member privileges
PROHIBITED Materials
- Any sample that will outgas under vacuum (check with staff if uncertain). This includes but is not limited to samples containing elemental sulpher or phosphorus
- Samples containing any degree of moisture
- Geological samples
- Biological samples
- Samples containing dust or powders, including magnetic powders or nanoparticles.
NOTE: All samples must be suitable for a cleanroom environment. Please note that other excellent SEM capabilities exist on campus including in WATLAB.
Authorized Substrates & Thin Films
- Polymer films: All photo/e-beam resists, conditional on them being fully baked/cured based on manufacturers instructions.
- CNTs, graphene
- Silicon
- III-V's
- ZnSe
- Ag
- Al
- Al2O3
- Au
- Bi2Te3
- Cr
- Cu
- EuS
- Ge
- HfO2
- ITO
- Nb
- Ni
- Pd
- Pt
- SiNx
- SiO2
- Ti
- TiO2
- TiW
- W
NOTE:
- Copper or Carbon tape should be used to adhere all samples loaded into the system. Silver-epoxy can be used if allowed to dry in air for 10 minutes prior to loading into the system
- If any other materials are present on your samples, CONTACT STAFF BEFORE PROCEEDING